- Description
- Attributes
- Documents
- 4.3-inch color touch screen
- 4-wire Kelvin test leads included
- Temperature measurement with correction
- Low power resistance mode to protect DUT
- Manual or auto ranging
- Adjustable measurement speed for fast readout or better accuracy
- Fast measurement speed up to 20 ms/reading to increase manufacturing throughput
- 5-1/2 digit display*
- Offset voltage compensation (OVC)
- Cable compensation (0 ADJ)
- BIN comparator function to sort components in up to 10 bin locations
- Bin-sorting with statistical measurement
- Store/recall 30 instrument settings
- Screen capture to USB drive
- Selectable power line filter
- Trigger delay
- Handler interface for easy integration with a component handler
- Standard RS232, USB (USBTMC and virtual COM), and LAN interfaces
- Three-year warranty
Quickly sort components using the 10 bins of the 2841. The results can be displayed on the screen or output via the handler interface. High and low limits can be set up in absolute or tolerance mode. |
High accuracy and resolution measurements in the milliohm range
B&K's 2841 DC resistance meter is ideally suited for measuring contact resistance of relays, switches, and PCB traces, typically outside the resistance range of multimeters. The 2841 is equipped with an intuitive front panel complete with a keypad and a vivid 4.3-inch color touch screen. Additionally, it supports fast cycle times for production automation and includes a BIN comparator function.
Intuitive user interface
The touch screen and keys both provide a convenient interface for setting parameters quickly and precisely. Both models also support one touch zoom to enhance the readability of displayed measurements.
Extended range, accuracy, temperature measurement and functions
The 2841 allows measurement of low resistivity materials with offset voltage compensation to reduce the influence of thermal EMF. Cable error detection displays a message when one of the terminals is open, which helps identify defective test leads or poor contact that can cause unreliable measurements.
LPR (Lower Power Resistance) measurement
The 2841 features low power resistance testing (LPR) modes and variable measurement speeds that are suited for a wide range of applications. The handler interface with high-speed measurement capabilities enable the 2841 to evaluate a large quantity of components and be integrated into an automated test system.
Temperature measurement with correction
The 2841 includes two temperature compensation functions: correction (TC) and conversion (Δt). The TC function corrects for ambient temperature changes that cause different resistance measurements of the same component. The temperature conversion (Δt) function can be used to evaluate a coil’s resistance before and after operation to calculate a temperature change. These additional features make the 2841 ideal for evaluating coils, motor windings, transformers, actuators and conductive materials.
Offset voltage compensation (OVC)
OVC switches the polarity and averages the forward and reverse polarity readings. This enables accurate measurements by minimizing the adverse effects of thermal EMFs or small biases in a circuit (e.g., from a capacitor's dielectric absorption when measuring a PCB trace).
Kelvin Test Leads included Measure the resistance of motor and transformer windings, relay and switch contacts, and other conductive materials and components. |
Model | 2840 | 2841 |
---|---|---|
Display Range | 1 μΩ to 20 kΩ | 0.1 μΩ to 100 MΩ |
Basic Accuracy | 0.1% | 0.01% |
Measurement Resolution | 1 μΩ | 0.1 μΩ |
Displayed Digits | 4-1/2 | 5-1/2 |
Functions | R and LPR | R, R-T, T, LPR, LPR-T |
Resistance Measurement Ranges | 7 + Auto | 11 + Auto |
Temperature Compensation | n/a | √ |
Number of Results Bins | 3 | 10 |
Included with 2841: AC power cord, user's manual, TLDK1 kelvin test leads, TPTC2 temperature probe, and test report and certificate of calibration.