- Description
- Attributes
- Documents
Tektronix P6021A AC Current Oscilloscope Probe
Features and Specifications
- 120 Hz to 60 MHz
- 10.6 A RMS, 250 A peak, 10 mA sensitivity
- For 1 MΩ inputs
- Shielded probe head
- AC only
- Split core construction allows easy circuit connection
- 1.5 m (5 ft) cable
The P6021A Current Probe provides versatile AC current measurements. These probes provide accurate current measurements over a wide range of frequencies. They allow current measurements without breaking the circuit by clipping onto the current carrying conductor. Shielded probe heads are not grounded when the slides are in their open positions, eliminating accidental grounding of the circuit under test. Additionally, the P6021A is certified for bare wire voltages up to 150 V CAT II.
For general purpose applications, the P6021A provides wide-band performance with excellent low-frequency characteristics. Bandwidth is 120 Hz to 60 MHz. The passive termination is switchable from 2 mA/mV to10 mA/mV.