- RF generator for conducted and magnetic field tests
- Simple expansion with external amplifier via 2nd generator output
- SCPI command set enables easy integration into own software systems
- Temperature measuring input, e.g. for monitoring and displaying the BCI clamp temperature
- Input for external pulse modulation
- Configurable, digital 8-channel user port
- RF generator for conducted and magnetic field tests
- 100 kHz - 400 MHz, amplifier 75 W
- Maximum test level: 30 V (40 V) with 80% AM (without 6 dB)
- Built-in directional coupler with software HELIA 7 - Basic
- Interfaces: USB, LAN, GPIB (option)
- Warranty: 3 years
The CDG 7000 is a new test generator for all standards for immunity to conducted disturbances and magnetic fields induced by radio-frequency fields, such as IEC 61000-4-6 - including BCI tests (ISO 11452-4). The CDG 7000 generates interferences as defined in IEC / EN 61000-4-6 - immunity to conducted disturbances induced by radio-frequency fields. The standard describes a test setup in which these high-frequency interferences can be influenced on a EUT without a complicated structure with antennas, field instrumentation and shielded rooms. By using coupling networks and coupling clamp's sine waves are induced directly into power and signal lines. The test object retains its original place in the device structure, so that the system can be tested in its overall function.
The versatile combination device includes an RF signal generator, an RF power amplifier, a 3-channel RF voltmeter and a directional coupler.